Identificação de microcistina LR ao nível molecular empregando microscopia de força atômica


Autoria(s): Etchegaray,Augusto; Bueno,Carolina de Castro; Teschke,Omar
Data(s)

01/01/2010

Resumo

Microcystins are non-ribosomal peptides that must be detected for its health concern. Here, microcystin LR and its specific antibody were respectively tethered to the substrate and to the tip of an atomic force microscope, after surface functionalization using 3-aminopropyltriethoxysilane and glutaraldehyde. Functionalization was confirmed comparing topographic images taken on bare and modified tips. Force versus distance curves were successfully used to measure the specific antibody-antigen interactions comparing with a control in which microcystin was initially blocked by incubation with free antibodies. The results showed unequivocally the specific recognition of MLR, suggesting that this method could be useful for biosensor development.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422010000900004

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.33 n.9 2010

Palavras-Chave #microcystin LR #atomic force microscope #biosensor development