Validação de método para determinação de resíduos de amoxicilina aplicado à validação de limpeza em indústria farmacêutica de penicilânicos
| Data(s) |
01/01/2010
|
|---|---|
| Resumo |
The aim of this work was the single-laboratory validation of a quantitative method for the determination of amoxicillin residues in support of cleaning control and validation. Linearity was demonstrated between 2.5 and 17.5 μg/mL, without matrix effects. Mean recoveries ranged from 84.00 to 103.74% and the relative standard deviation under repetitivity and within-reproducibility conditions were from 0.58 to 4.20% and from 0.79 to 4.39%, respectively. The theoretical limits of detection and quantification were 0.133 and 0.442 μg/mL, respectively. The studied method was suitable for cleaning control purpose within good manufacturing practices. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422010000400038 |
| Idioma(s) |
pt |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.33 n.4 2010 |
| Palavras-Chave | #amoxicillin #method validation #cleaning validation |
| Tipo |
journal article |