Otimização da análise isotópica de UF6 utilizando-se a técnica de espectrometria de massas por quadrupolo


Autoria(s): Porto,Peterson; Bustillos,José Oscar Vega
Data(s)

01/01/2008

Resumo

A procedure for determining of the isotope ratio 235U/238U in UF6 samples was established using a quadrupole mass spectrometer with ionization by electron impact. The following items were optimized in the spectrometer: the parameters in the ion source that provided the most intense peak, with good shape, for the most abundant isotope; the resolution that reduced the non linear effects and the number of analytical cycles that reduced the uncertainty in the results. The measurement process was characterized with respect to the effects of mass discrimination, linearity and memory effect.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422008000600047

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.31 n.6 2008

Palavras-Chave #mass spectrometry #isotopic analysis #uranium hexafluoride
Tipo

journal article