Quantificação simultânea de indicadores biológicos de exposição a solventes orgânicos por cromatografia líquida de alta eficiência


Autoria(s): Bulcão,Rachel; Maria,Lucas Santa; Charão,Mariele; Moro,Angela; Roehrs,Miguel; Garcia,Solange Cristina; Limberger,Renata Pereira
Data(s)

01/01/2008

Resumo

Biological monitoring is very important to guarantee health to workers. This method was developed for simultaneous determination of xylene, toluene, styrene and ethylbenzene metabolites. It involves only dilution and centrifugation of urine samples and improved chromatographic conditions. Analyses show recovery > 95%; r² > 0.99; intermediate precision CV% < 6% and % bias < ±10. Exposed subjects presented at least three metabolites in urine. The method proved to be feasible, reliable and important in biological monitoring, especially in exposure to organic solvent mixtures.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422008000600013

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.31 n.6 2008

Palavras-Chave #HPLC #organic solvents #biological exposure index
Tipo

journal article