Determination of etoricoxib in human plasma using automated on-line solid-phase extraction coupled with LC-APCI/MS/MS


Autoria(s): Dalmora,Sérgio Luiz; Brum Junior,Liberato; Ferretto,Ricardo Machado; Oliveira,Paulo Renato de; Barth,Thiago; Sangoi,Maximiliano da Silva
Data(s)

01/01/2008

Resumo

A liquid chromatography-tandem mass spectrometry method with atmospheric pressure chemical ionization (LC-APCI/MS/MS) was validated for the determination of etoricoxib in human plasma using antipyrin as internal standard, followed by on-line solid-phase extraction. The method was performed on a Luna C18 column and the mobile phase consisted of acetonitrile:water (95:5, v/v)/ammonium acetate (pH 4.0; 10 mM), run at a flow rate of 0.6 mL/min. The method was linear in the range of 1-5000 ng/mL (r²>0.99). The lower limit of quantitation was 1 ng/mL. The recoveries were within 93.72-96.18%. Moreover, method validation demonstrated acceptable results for the precision, accuracy and stability studies.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422008000300021

Idioma(s)

en

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.31 n.3 2008

Palavras-Chave #etoricoxib #LC-APCI/MS/MS #on-line solid-phase extraction
Tipo