Validação de método por cromatografia líquida de alta eficiência para determinação da lamivudina e zidovudina em comprimidos
Data(s) |
01/10/2007
|
---|---|
Resumo |
An HPLC method was validated to assay lamivudine and zidovudine combined in tablets. The chromatographic separation was carried out using methanol and acetate buffer pH 6.5 (50:50 v/v) and a RP-18 column, as mobile and stationary phase, respectively. The UV detection was at 270 nm. The method was linear in the range of 24 - 36 µg/mL (lamivudine) and 48 - 72 µg/mL (zidovudine). The recovery (accuracy) ranged from 101.35% to 103.04% and the precision (repeatability and intermediate precision) was less than 2%. The method can be also applied to the quantification of these drugs in the dissolution test of tablets containing both drugs. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422007000500033 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Fonte |
Química Nova v.30 n.5 2007 |
Palavras-Chave | #HPLC #lamivudine #zidovudine |
Tipo |
journal article |