Digestão de óleo lubrificante encapsulado em forno de microondas com radiação focalizada por adição de amostra ao reagente pré-aquecido


Autoria(s): Bressani,Flavia A.; Silva,Hamilton O.; Nóbrega,Joaquim A.; Costa,Letícia M.; Nogueira,Ana Rita A.
Data(s)

01/12/2006

Resumo

The applicability of the recently proposed procedure based on gradual sample addition to microwave-assisted pre-heated concentrated acid is limited by the sample viscosity. In this work, lubricating oil samples with high viscosity were encapsulated and manually added to the microwave-assisted pre-heated concentrated digestion mixture. The procedure was applied for determination of Al, Ca, Cr, Cu, Fe, Mg, Ni, P, Pb, Si, Sn, Sr, V, W, and Zn in lubricating oil by inductively coupled plasma optical emission spectrometry (ICP OES). Determined and certified values for Ca, Mg, P, and Zn in lubricating oil were in agreement at a 95% confidence level.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000600012

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.29 n.6 2006

Palavras-Chave #encapsulated lubricating oil #sample addition #focused-microwave
Tipo

journal article