Construção e aplicação de biossensores usando diferentes procedimentos de imobilização da peroxidase de vegetal em matriz de quitosana
Data(s) |
01/10/2006
|
---|---|
Resumo |
Biosensors were developed by immobilization of gilo (Solanum gilo) enzymatic extract on chitosan biopolymers using three different procedures: glutaraldehyde, carbodiimide/glutaraldehyde and epichlorohydrin/glutaraldehyde. The best biosensor performance was obtained after the immobilization of peroxidase on chitosan with epichlorohydrin/glutaraldehyde. Linear analytical curves for hydroquinone concentrations from 2.5x10-4 to 4.5x10-3 mol L-1 with a detection limit of 2.0x10-6 mol L-1 and recovery of hydroquinone ranging from 95.1 to 105% were obtained. The relative standard deviation was < 1.0 % for a solution of 3.0x10-4 mol L-1 hydroquinone and 2.0x10-3 mol L-1 hydrogen peroxide in 0.1 mol L-1 phosphate buffer solution at pH 7.0 (n=8). The lifetime of this biosensor was 6 months (at least 300 determinations). |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422006000500009 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Fonte |
Química Nova v.29 n.5 2006 |
Palavras-Chave | #chitosan #biosensor #peroxidase |
Tipo |
journal article |