Resumo

Biosensors were developed by immobilization of gilo (Solanum gilo) enzymatic extract on chitosan biopolymers using three different procedures: glutaraldehyde, carbodiimide/glutaraldehyde and epichlorohydrin/glutaraldehyde. The best biosensor performance was obtained after the immobilization of peroxidase on chitosan with epichlorohydrin/glutaraldehyde. Linear analytical curves for hydroquinone concentrations from 2.5x10-4 to 4.5x10-3 mol L-1 with a detection limit of 2.0x10-6 mol L-1 and recovery of hydroquinone ranging from 95.1 to 105% were obtained. The relative standard deviation was < 1.0 % for a solution of 3.0x10-4 mol L-1 hydroquinone and 2.0x10-3 mol L-1 hydrogen peroxide in 0.1 mol L-1 phosphate buffer solution at pH 7.0 (n=8). The lifetime of this biosensor was 6 months (at least 300 determinations).