Determinação de arsênio em águas contaminadas usando fluorescência de raios-X por energia dispersiva


Autoria(s): Pataca,Luiz Carlos M.; Bortoleto,Gisele G.; Bueno,Maria Izabel M. S.
Data(s)

01/08/2005

Resumo

This work proposes a simple, fast and inexpensive method to determine As in natural waters, using X-ray fluorescence. 50 µL of each sample containing 100 mg L-1 of yttrium as internal standard were deposited over a 2.5 µm thickness MylarTM film. The samples were dried at 50 °C for 2 h. X-ray spectra were obtained using an EDXRF apparatus. The accuracy was determined by analyte addition/recovery and by comparison with Hydride Generation Atomic Absorption Spectrometry (HG AAS). A recovery of about 100% was obtained and the results were in good agreement with HG AAS. The method showed a relative standard deviation of 6.8% and a detection limit of 10.5 µg L-1 of As.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422005000400004

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.28 n.4 2005

Palavras-Chave #arsenic #natural water #EDXRF
Tipo

journal article