Imobilização de enzimas a partir de "kit" comercial: determinação de parâmetros metabólicos em sangue animal empregando multicomutação em fluxo
Data(s) |
01/06/2005
|
---|---|
Resumo |
Automatic flow procedures based on the multicommutation concept, dedicated to the determination of 3-hydroxybutyrate, glucose and cholesterol are proposed. The enzymes were immobilized on glass beads and packed into mini-columns that were coupled to a flow system. Sampling throughputs of 55, 40 and 40 determinations per hour, linear response from 10 to 150, 50 to 600, 25 to 125 mg L-1, detection limits of 1.5, 14 and 4 mg L-1 and relative standard deviations of 1, 2 and 2% for 3-hydroxybutyrate, glucose and cholesterol, respectively, were achieved. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422005000300010 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Fonte |
Química Nova v.28 n.3 2005 |
Palavras-Chave | #enzymatic reactions #flow injection analysis #commercial kit |
Tipo |