Acoplamento de cela de difusão gasosa a sistema de análise por injeção seqüencial visando a determinação espectrofotométrica de sulfeto


Autoria(s): Silva,Marcelo Solitrenick Pinto; Masini,Jorge Cesar
Data(s)

01/08/2004

Resumo

A new configuration for coupling a gas diffusion cell to a sequential injection system is presented. The matrix exchange is made without the need for additional rotary injection valves or peristaltic pumps, keeping the original mechanical components of the sequential injection apparatus: one syringe pump (or peristaltic pump) and one selection valve. The system was tested constructing analytical curves for sulfide exploring the formation of the methylene blue dye. The proposed method has a detection limit of 60 µg L-1 S2-, with a linear dynamic range between 0.10 and 4.0 mg L-1 S2- concentrations, with a sampling frequency of 20 h-1.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422004000400006

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.27 n.4 2004

Palavras-Chave #sequential injection analysis #gas diffusion #sulfide
Tipo

journal article