Determinação enzimática de dopamina em formulações farmacêuticas utilizando sistema de análise por injeção em fluxo com extrato bruto de abacate (Persea americana)
| Data(s) |
01/03/2003
|
|---|---|
| Resumo |
In this work, a spectrophotometric flow injection analysis system using a crude extract of avocado (Persea americana) as a source of polyphenol oxidase to dopamine determination was developed. The substrates and enzyme concentrations from 2.4x10-7 to 5.3x10-4 mol L-1 and 28 to 332 units mL-1 were evaluated, respectively. In addition, the FIA parameters such as sample loop (50 to 500 µL), flow rate (1.4 to 4.3 mL min-1) and reactor length (100 to 500 cm) were also evaluated in a 0.1 mol L-1 phosphate buffer solution (pH 7.0). Dopamine solution concentrations were determined using 277 units mL-1 enzyme solution, 400 mL enzyme loop, 375 µL sample loop, 2.2 mL min-1 flow rate and a reactor of 350 cm. The analytical curve showed a linearity from 5.3x10-5 to 5.3x10-4 mol L-1 dopamine with a detection limit of 1.3x10-5 mol L-1. The analytical frequency was 46 h-1 and the RSD lower than 0.5% for 5.3x10-4 mol L-1 dopamine solution (n=10). A paired t-test showed that all results obtained for dopamine in commercial formulations using the proposed flow injection procedure and a spectrophotometric procedure agree at the 95% confidence level. |
| Formato |
text/html |
| Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40422003000200010 |
| Idioma(s) |
pt |
| Publicador |
Sociedade Brasileira de Química |
| Fonte |
Química Nova v.26 n.2 2003 |
| Palavras-Chave | #dopamine #flow injection analysis #crude extract |
| Tipo |
journal article |