Titulador potenciométrico automatizado baseado em sistema de fluxo monossegmentado


Autoria(s): Ganzarolli,Edgard Moreira; Lehmkuhl,Arilson; Queiróz,Roldão Rosevelt R. de; Souza,Ivan Gonçalves de
Data(s)

01/02/1999

Resumo

A new automated system for acid-base flow titrations is proposed. In the operation mode, several sample to titrant volumetric ratios are injected in an air segmented plug. Five three way solenoid valves and three acrilic junctions, assembled in a hidrodynamic injection system, were accountable for the monosegmented reagents plug formation. A turbulent flow reactor was used for a perfect mix of reagents in the plug. The detector system employed a glass combined electrode fitted in an acrilic holder. Titrations of hydrochloric, nitric and acetic acids, in several concentrations, were performed with standard sodium hidroxide, for evaluation of the efficiency of the system. The relative standard deviation of the determinations was about ±0,5% and each titration was carried out in 3-4 minutes. A Quick BASIC 4.5® program was developed for the titrator control.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421999000100011

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.22 n.1 1999

Palavras-Chave #automation #potentiometric titrations #flow analysis
Tipo

journal article