Sistema automático para análise direta de ligas metálicas por eletrodissolução anódica e espectrometria de absorção atômica
Data(s) |
01/02/1999
|
---|---|
Resumo |
An automatic system for the direct determination of lead and tin by atomic absorption spectrometry is described. The on-line treatment of the metallic samples was obtained by anodic electrodissolution in a flow injection system. Lead was determined by flame atomic absorption spectrometry (FAAS) and tin by graphite furnace atomic absorption spectrometry (GFAAS). A computer program managed the current source and the solenoid valves that direct the fluids. Good linear correlations between absorbance and current intensity for lead and tin were observed. Results were in agreement with the certified values. Precision was always better than 5%. The recommended procedure allows the direct determination of 60 or 30 elements/h using FAAS or GFAAS, respectively. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421999000100004 |
Idioma(s) |
pt |
Publicador |
Sociedade Brasileira de Química |
Fonte |
Química Nova v.22 n.1 1999 |
Palavras-Chave | #anodic electrodissolution #flow injection analysis #atomic absoption spectrometry |
Tipo |