Sistema automático para análise direta de ligas metálicas por eletrodissolução anódica e espectrometria de absorção atômica


Autoria(s): Silva,José Bento B.; Giacomelli,Maria B. O.; Lehmkuhl,Arilson; Curtius,Adilson J.; Queiróz,Roldão R. U.; Souza,Ivan G. de
Data(s)

01/02/1999

Resumo

An automatic system for the direct determination of lead and tin by atomic absorption spectrometry is described. The on-line treatment of the metallic samples was obtained by anodic electrodissolution in a flow injection system. Lead was determined by flame atomic absorption spectrometry (FAAS) and tin by graphite furnace atomic absorption spectrometry (GFAAS). A computer program managed the current source and the solenoid valves that direct the fluids. Good linear correlations between absorbance and current intensity for lead and tin were observed. Results were in agreement with the certified values. Precision was always better than 5%. The recommended procedure allows the direct determination of 60 or 30 elements/h using FAAS or GFAAS, respectively.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421999000100004

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.22 n.1 1999

Palavras-Chave #anodic electrodissolution #flow injection analysis #atomic absoption spectrometry
Tipo

journal article