Uma célula simples para adaptação de eletrodos seletivos comerciais em sistemas de análises em fluxo


Autoria(s): Marin,Maria Angélica Bonadiman; Ganzarolli,Edgard Moreira; Queiróz,Roldão Rooseveld Urzêdo de; Souza,Ivan Gonçalves de
Data(s)

01/11/1998

Resumo

A simple flow cell for potentiometric detection is described. It was assembled by making use of two perspex pieces fixed together by means of four screws, and allow the connection of plane membrane conventional electrodes to flow system. Details about its construction are presented. The device performance was evaluated by making use of a cyanide ion-selective electrode. The relative standard deviation was about 0.5% with a detection limit of 8.0 x 10-6 mol CN- dm-3. Under experimental conditions, the linear range was 10-5 to10-2 mol dm-3.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421998000600022

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.21 n.6 1998

Palavras-Chave #flow cell #potentiometric detector #continuous flow analysis