Utilização de uma interface do tipo "particle beam" para a obtenção de espectros de massas de compostos pouco voláteis em solução


Autoria(s): Augusti,Rodinei; Dias,Adelson de O.; Fortes,Isabel C. P.
Data(s)

01/10/1998

Resumo

Due to source contamination and wearing of instrument components problems caused by the direct insertion probe technique, a new way of introduction of low volatile compounds into mass spectrometer was tested. This new scheme comprises the introduction of the low volatile compounds solutions via a six port valve connected to a particle beam interface. Solutions of isatin were injected into this system and the best results were obtained with CH2Cl2, CH3OH and CH3CN. The solution inlet system has shown to be advantageous over the conventional way of direct insertion probe introduction.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421998000500019

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.21 n.5 1998

Palavras-Chave #mass spectrometry #particle beam interface #low volatile compounds
Tipo

journal article