Determinações multi-elementares de metais, Sulfato e Cloreto em amostras de águas por espectrometria de massas com fonte de plasma (ICP-MS)


Autoria(s): Menegário,A. A.; Giné,M. F.
Data(s)

01/07/1998

Resumo

The multi-element determination of Al, Cr, Mn, Ni, Cu, Zn, Cd, Ba, Pb, SO4= and Cl- in riverine water samples was accomplished by inductively coupled plasma mass spectrometry (ICP-MS). The sample passed through a column containing the anionic resin AG1-X8 and the metals were determined directly. The retained anionic species were eluted and SO4= and Cl- were determined at m/z 48 and 35 correspondent to the ions SO+ and Cl+ formed at the plasma. Accuracy for metals was assessed by analysing the certified reference TM-26 (National Water Research Institute of Canada). Results for SO4= and Cl- were in agreement with those obtained by turbidimetry and spectrophotometry. LOD's of 0.1 µg l-1 for Cd, Ba and Pb; 0.2 µg l-1 for Al, Mn and Cu; 0.5 µg l-1 for Cr; 0.9 for Zn; 2.0 µg l-1for Ni , 60 µg l-1 for S and 200 µg l-1 Cl were attained.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-40421998000400007

Idioma(s)

pt

Publicador

Sociedade Brasileira de Química

Fonte

Química Nova v.21 n.4 1998

Palavras-Chave #water analysis #inductively coupled plasma mass spectrometry #anion exchange #flow system
Tipo

journal article