Monte Carlo derivation of filtered tungsten anode X-ray spectra for dose computation in digital mammography
Data(s) |
01/12/2015
|
---|---|
Resumo |
Abstract Objective: Derive filtered tungsten X-ray spectra used in digital mammography systems by means of Monte Carlo simulations. Materials and Methods: Filtered spectra for rhodium filter were obtained for tube potentials between 26 and 32 kV. The half-value layer (HVL) of simulated filtered spectra were compared with those obtained experimentally with a solid state detector Unfors model 8202031-H Xi R/F & MAM Detector Platinum and 8201023-C Xi Base unit Platinum Plus w mAs in a Hologic Selenia Dimensions system using a direct radiography mode. Results: Calculated HVL values showed good agreement as compared with those obtained experimentally. The greatest relative difference between the Monte Carlo calculated HVL values and experimental HVL values was 4%. Conclusion: The results show that the filtered tungsten anode X-ray spectra and the EGSnrc Monte Carlo code can be used for mean glandular dose determination in mammography. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842015000600006 |
Idioma(s) |
en |
Publicador |
Colégio Brasileiro de Radiologia e Diagnóstico por Imagem |
Fonte |
Radiologia Brasileira v.48 n.6 2015 |
Palavras-Chave | #Mammography #X-ray spectra #HVL #Monte Carlo |
Tipo |