Comparison of standardized uptake values measured on 18F-NaF PET/CT scans using three different tube current intensities


Autoria(s): Valadares,Agnes Araujo; Duarte,Paulo Schiavom; Woellner,Eduardo Bechtloff; Coura-Filho,George Barberio; Sapienza,Marcelo Tatit; Buchpiguel,Carlos Alberto
Data(s)

01/02/2015

Resumo

Objective: To analyze standardized uptake values (SUVs) using three different tube current intensities for attenuation correction on 18FNaF PET/CT scans. Materials and Methods: A total of 254 18F-NaF PET/CT studies were analyzed using 10, 20 and 30 mAs. The SUVs were calculated in volumes of interest (VOIs) drawn on three skeletal regions, namely, right proximal humeral diaphysis (RH), right proximal femoral diaphysis (RF), and first lumbar vertebra (LV1) in a total of 712 VOIs. The analyses covered 675 regions classified as normal (236 RH, 232 RF, and 207 LV1). Results: Mean SUV for each skeletal region was 3.8, 5.4 and 14.4 for RH, RF, and LV1, respectively. As the studies were grouped according to mAs value, the mean SUV values were 3.8, 3.9 and 3.7 for 10, 20 and 30 mAs, respectively, in the RH region; 5.4, 5.5 and 5.4 for 10, 20 and 30 mAs, respectively, in the RF region; 13.8, 14.9 and 14.5 for 10, 20 and 30 mAs, respectively, in the LV1 region. Conclusion: The three tube current values yielded similar results for SUV calculation.

Formato

text/html

Identificador

http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0100-39842015000100007

Idioma(s)

en

Publicador

Colégio Brasileiro de Radiologia e Diagnóstico por Imagem

Fonte

Radiologia Brasileira v.48 n.1 2015

Palavras-Chave #18F-NaF PET/CT #SUV #Tube current #mAs #18F-NaF PET/CT #SUV #Corrente de tubo #mAs
Tipo

journal article