Use of a ventricular septal defect occluder for apical closure in transapical aortic valve replacement.


Autoria(s): Ferrari E.; Locca D.; Berdajs D.; Marcucci C.; Gronchi F.; Lavanchy J.; Prêtre R.; Tozzi P.
Data(s)

2015

Resumo

During transapical transcatheter aortic valve replacement (TA-TAVR), the apical closure remains a challenge for the surgeon, having the risk for ventricular tear and massive bleeding. Apical closure devices are already under clinical evaluation, but only a few can lead to a full percutaneous TA-TAVR. We describe the successful use of a 9-mm myocardial occluder (ventricular septal defect occluder) that was used to seal the apex after a standard TA-TAVR (using the Sapien XT 23-mm transcatheter valve and the Ascendra + delivery system). The placement of the nonmodified myocardial occluder was performed through the Ascendra + delivery system, with a very small amount of blood loss and an acceptable sealing of the apical tear. This approach is feasible and represents a further step toward true-percutaneous transapical heart valve procedures. Modified apical occluders are under evaluation in animal models.

Identificador

https://serval.unil.ch/?id=serval:BIB_07EDA7E358FB

isbn:1559-0879 (Electronic)

pmid:25621876

doi:10.1097/IMI.0000000000000118

Idioma(s)

en

Fonte

Innovations, vol. 10, no. 1, pp. 68-70

Tipo

info:eu-repo/semantics/article

article