Precision of the Retinal Tomograph as a Screening Device.


Autoria(s): Bergin C.; Oleszczuk J.; Sharkawi E.
Data(s)

2015

Resumo

Purpose: To assess the diagnostic accuracy of the Heidelberg Retinal Tomograph 3 (HRT3) as a screening device in comparison with the reference standard of Octopus standard automated perimetry results (SAP) combined with clinical findings. Methods: All patients underwent screening examinations and investigations within a single day. Abnormal screening results were classified as follows: The HRT3: Either "borderline" or "outside normal limits" using the global Moorfields classification (MFC); SAP and clinical exam: A mean defect > 2.4 dB or "outside normal limits" clear text analysis of SAP; and one of the following i) IOP > 21 mmHg, ii) Van Herrick < ¼, iii) cup disc ratio > 0.55, iv) optic nerve head abnormality, v) narrow iridocorneal angle or vi) evidence of peripheral anterior synechiae on gonioscopy. Results: The mean age of the participants was 59.9 years (± 14.8 [21, 91]). Twenty-three subjects (16 %) were classified as abnormal on SAP and clinical exam. The HRT3 classification had a sensitivity of 30 % (95 % CI [16 %, 51 %]) with associated specificity of 58 % (95 % CI [49 %, 66 %]). Of the sixty subjects classified as borderline or outside normal limits with the HRT MFC global result, seven subjects were also abnormal according to SAP and clinical exam. Conclusion: The results suggest that the HRT3 may not be suitable as a sole screening device; however, further investigation is necessary.

Identificador

https://serval.unil.ch/?id=serval:BIB_41A3D5F08D17

isbn:1439-3999 (Electronic)

pmid:25902093

doi:10.1055/s-0035-1545814

isiid:000353501000026

Idioma(s)

en

Fonte

Klinische Monatsblatter Fur Augenheilkunde, vol. 232, no. 4, pp. 438-441

Tipo

info:eu-repo/semantics/article

article