Minimization of Nyquist ghosting for echo-planar imaging at ultra-high fields based on a "negative readout gradient" strategy.


Autoria(s): van der Zwaag W.; Marques J.P.; Lei H.; Just N.; Kober T.; Gruetter R.
Data(s)

2009

Resumo

PURPOSE: To improve the traditional Nyquist ghost correction approach in echo planar imaging (EPI) at high fields, via schemes based on the reversal of the EPI readout gradient polarity for every other volume throughout a functional magnetic resonance imaging (fMRI) acquisition train. MATERIALS AND METHODS: An EPI sequence in which the readout gradient was inverted every other volume was implemented on two ultrahigh-field systems. Phantom images and fMRI data were acquired to evaluate ghost intensities and the presence of false-positive blood oxygenation level-dependent (BOLD) signal with and without ghost correction. Three different algorithms for ghost correction of alternating readout EPI were compared. RESULTS: Irrespective of the chosen processing approach, ghosting was significantly reduced (up to 70% lower intensity) in both rat brain images acquired on a 9.4T animal scanner and human brain images acquired at 7T, resulting in a reduction of sources of false-positive activation in fMRI data. CONCLUSION: It is concluded that at high B(0) fields, substantial gains in Nyquist ghost correction of echo planar time series are possible by alternating the readout gradient every other volume.

Identificador

http://serval.unil.ch/?id=serval:BIB_EC8D61037BB7

isbn:1522-2586[electronic]

pmid:19856451

doi:10.1002/jmri.21951

isiid:00027124790003

Idioma(s)

en

Fonte

Journal of Magnetic Resonance Imaging, vol. 30, no. 5, pp. 1171-1178

Palavras-Chave #Nyquist ghost; Ghost correction; EPI; FMRI; Nyquist ghost; ghost correction; EPI; FMRI
Tipo

info:eu-repo/semantics/article

article