Short-term stability of testosterone and epitestosterone conjugates in urine samples: quantification by liquid chromatography-linear ion trap mass spectrometry.
Data(s) |
2006
|
---|---|
Resumo |
A simple method using liquid chromatography-linear ion trap mass spectrometry for simultaneous determination of testosterone glucuronide (TG), testosterone sulfate (TS), epitestosterone glucuronide (EG) and epitestosterone sulfate (ES) in urine samples was developed. For validation purposes, a urine containing no detectable amount of TG, TS and EG was selected and fortified with steroid conjugate standards. Quantification was performed using deuterated testosterone conjugates to correct for ion suppression/enhancement during ESI. Assay validation was performed in terms of lower limit of detection (1-3ng/mL), recovery (89-101%), intraday precision (2.0-6.8%), interday precision (3.4-9.6%) and accuracy (101-103%). Application of the method to short-term stability testing of urine samples at temperature ranging from 4 to 37 degrees C during a time-storage of a week lead to the conclusion that addition of sodium azide (10mg/mL) is required for preservation of the analytes. |
Identificador |
http://serval.unil.ch/?id=serval:BIB_E03E9E4FCE27 isbn:1570-0232[print], 1570-0232[linking] pmid:16891159 doi:10.1016/j.jchromb.2006.07.010 isiid:000242328300024 |
Idioma(s) |
en |
Fonte |
Journal of Chromatography. B, Analytical Technologies In the Biomedical and Life Sciences, vol. 844, no. 1, pp. 168-174 |
Palavras-Chave | #Calibration; Child, Preschool; Epitestosterone/urine; Female; Humans; Mass Spectrometry/methods; Sensitivity and Specificity; Testosterone/urine |
Tipo |