Performance of a chromatographic procedure used in the certification of reference material for polycyclic aromatic hydrocarbons in sewage sludge
Data(s) |
1995
|
---|---|
Resumo |
Semi-automatic capillary gas chromatographic method with classical flame ionization detection, which satisfies the conditions for required performance and gave acceptable results within the framework of an interlaboratory certification programme for PAHs in sewage sludge, is described. The interesting feature of the procedure is that it incorporates automatic operations such as sample fractionation by semi-preparative HPLC, fraction collection at signal level recognition and evaporation under nitrogen flow. Multiple injections in the GC capillary column are performed in the on-column mode via an autosampler with temperature-programmable injector. Automatic data acquisition and chromatogram treatment are made via computer software. This partially automatic procedure releases personnel from tedious and time-consuming tasks and its robust character was validated through the certification of reference material for PAHs in sewage sludge, demonstrating its reliable performance. |
Identificador |
http://serval.unil.ch/?id=serval:BIB_C7006C3D93C7 isbn:0026-3672 isiid:A1995TH37000023 |
Idioma(s) |
en |
Fonte |
Mikrochimica acta, vol. 120, no. 1-4, pp. 271-280 |
Tipo |
info:eu-repo/semantics/article article |