Validation of a wrist monitor for accurate estimation of RR intervals during sleep.
Data(s) |
2013
|
---|---|
Resumo |
While the incidence of sleep disorders is continuously increasing in western societies, there is a clear demand for technologies to asses sleep-related parameters in ambulatory scenarios. The present study introduces a novel concept of accurate sensor to measure RR intervals via the analysis of photo-plethysmographic signals recorded at the wrist. In a cohort of 26 subjects undergoing full night polysomnography, the wrist device provided RR interval estimates in agreement with RR intervals as measured from standard electrocardiographic time series. The study showed an overall agreement between both approaches of 0.05 ± 18 ms. The novel wrist sensor opens the door towards a new generation of comfortable and easy-to-use sleep monitors. |
Identificador |
http://serval.unil.ch/?id=serval:BIB_AEFC63AFA775 isbn:1557-170X (Print) pmid:24110980 doi:10.1109/EMBC.2013.6610793 |
Idioma(s) |
en |
Fonte |
Conference Proceedings : ... Annual International Conference of the IEEE Engineering in Medicine and Biology Society., vol. 2013, pp. 5493-5496 |
Tipo |
info:eu-repo/semantics/article article |