Polyimide/SU-8 catheter-tip MEMS gauge pressure sensor.


Autoria(s): Hasenkamp W.; Forchelet D.; Pataky K.; Villard J.; Lintel H.V.; Bertsch A.; Wang Q.; Renaud P.
Data(s)

2012

Resumo

This paper describes the development of a polyimide/SU-8 catheter-tip MEMS gauge pressure sensor. Finite element analysis was used to investigate critical parameters, impacting on the device design and sensing characteristics. The sensing element of the device was fabricated by polyimide-based micromachining on a flexible membrane, using embedded thin-film metallic wires as piezoresistive elements. A chamber containing this flexible membrane was sealed using an adapted SU-8 bonding technique. The device was evaluated experimentally and its overall performance compared with a commercial silicon-based pressure sensor. Furthermore, the device use was demonstrated by measuring blood pressure and heart rate in vivo.

Identificador

https://serval.unil.ch/?id=serval:BIB_A7ED28D9C197

isbn:1572-8781 (Electronic)

pmid:22639233

doi:10.1007/s10544-012-9661-8

isiid:000308963200003

http://my.unil.ch/serval/document/BIB_A7ED28D9C197.pdf

http://nbn-resolving.org/urn/resolver.pl?urn=urn:nbn:ch:serval-BIB_A7ED28D9C1973

Idioma(s)

en

Direitos

info:eu-repo/semantics/openAccess

Fonte

Biomedical Microdevices, vol. 14, no. 5, pp. 819-828

Tipo

info:eu-repo/semantics/article

article