Correction of through-plane deformation artifacts in stimulated echo acquisition mode cardiac imaging.


Autoria(s): Fahmy A.S.; Pan L.; Stuber M.; Osman N.F.
Data(s)

2006

Resumo

Attempts to use a stimulated echo acquisition mode (STEAM) in cardiac imaging are impeded by imaging artifacts that result in signal attenuation and nulling of the cardiac tissue. In this work, we present a method to reduce this artifact by acquiring two sets of stimulated echo images with two different demodulations. The resulting two images are combined to recover the signal loss and weighted to compensate for possible deformation-dependent intensity variation. Numerical simulations were used to validate the theory. Also, the proposed correction method was applied to in vivo imaging of normal volunteers (n = 6) and animal models with induced infarction (n = 3). The results show the ability of the method to recover the lost myocardial signal and generate artifact-free black-blood cardiac images.

Identificador

http://serval.unil.ch/?id=serval:BIB_9F5845A1EE74

isbn:0740-3194[print], 0740-3194[linking]

pmid:16408265

doi:10.1002/mrm.20781

isiid:000235326500023

Idioma(s)

en

Fonte

Magnetic Resonance In Medicine, vol. 55, no. 2, pp. 404-412

Palavras-Chave #Animals; Artifacts; Computer Simulation; Contrast Media; Dogs; Gadolinium DTPA; Heart/anatomy & histology; Humans; Image Enhancement/methods; Magnetic Resonance Imaging/methods; Myocardial Infarction/pathology
Tipo

info:eu-repo/semantics/article

article