Development, design and validation of solid reference samples.
Data(s) |
2014
|
---|---|
Resumo |
We developed a method of sample preparation using epoxy compound, which was validated in two steps. First, we studied the homogeneity within samples by scanning tubes filled with radioactive epoxy. We found within-sample homogeneity better than 2%. Then, we studied the homogeneity between samples during a 4.5 h dispensing time. The homogeneity between samples was found to be better than 2%. This study demonstrates that we have a validated method, which assures the traceability of epoxy samples. |
Identificador |
http://serval.unil.ch/?id=serval:BIB_9B4B5974E5E5 isbn:1872-9800 (Electronic) pmid:24378364 doi:10.1016/j.apradiso.2013.11.085 isiid:000335274000104 |
Idioma(s) |
en |
Fonte |
Applied Radiation and Isotopes, vol. 87, pp. 480-484 |
Tipo |
info:eu-repo/semantics/article article |