Validation of methane measurement using headspace-GC-MS and quantification by a stable isotope-labeled internal standard generated in situ.


Autoria(s): Varlet V.; Smith F.; Augsburger M.
Data(s)

2013

Resumo

A previous study has shown the possibility to identify methane (CH4 ) using headspace-GC-MS and quantify it with a stable isotope as internal standard. The main drawback of the GC-MS methods discussed in literature for CH4 measurement is the absence of a specific internal standard necessary to perform quantification. However, it becomes essential to develop a safer method to limit the manipulation of gaseous CH4 and to precisely control the injected amount of gas for spiking and calibration by comparison with external calibration. To avoid the manipulation of a stable isotope-labeled gas, we have chosen to generate a labeled gas as an internal standard in a vial on the basis of the formation of CH4 by the reaction of Grignard reagent methylmagnesium chloride with deuterated water. This method allows precise measurement of CH4 concentrations in gaseous sample as well as in a solid or a liquid sample after a thermodesorption step in a headspace vial. A full accuracy profile validation of this method is then presented.

Identificador

http://serval.unil.ch/?id=serval:BIB_79264189EEA6

isbn:1615-9314 (Electronic)

pmid:23585415

doi:10.1002/jssc.201300080

isiid:000320553800013

Idioma(s)

en

Fonte

Journal of Separation Science, vol. 36, no. 12, pp. 1967-1972

Palavras-Chave #Deuterated methane; methane; Stable isotope-labeled standard; volatile alkane
Tipo

info:eu-repo/semantics/article

article