Controlled dynamic generation of standard 1,6-hexamethylene diisocyanate aerosols


Autoria(s): Huynh Cong Khanh; Vu Duc Trinh; Savolainen Heikki
Data(s)

1989

Resumo

A procedure for the dynamic generation of 1,6-hexamethylene diisocyanate (HDI) aerosol atmospheres of 70 micrograms m-3 (0.01 ppm) to 1.75 mg m-3 (0.25 ppm), based on the precise control of the evaporation of pure liquid HDI and subsequent dilution with air, was developed. The apparatus consisted of a home-made glass nebulizer coupled with a separation stage to exclude non-respirable droplets (greater than 10 microns). The aerosol concentrations were achieved by passing air through the nebulizer at 1.5-4.5 l. min-1 to generate dynamically 0.01-0.25 ppm of diisocyanate in an experimental chamber of 8.55 m3. The distribution of the liquid aerosol was established with an optical counter and the diisocyanate concentration was determined from samples collected in impingers by a high-pressure liquid chromatographic method. The atmospheres generated were suitable for the evaluation both of sampling procedures full scale, and of analytical methods: at 140 micrograms m-3 (0.02 ppm) they remained stable for 15-min provocation tests in clinical asthma, as verified by breath-zone sampling of exposed patients.

Identificador

http://serval.unil.ch/?id=serval:BIB_6908CEC7E128

isbn:0003-4878 (Print)

pmid:2604320

isiid:A1989CF38000007

Idioma(s)

en

Fonte

Annals of Occupational Hygiene, vol. 33, no. 4, pp. 573-581

Palavras-Chave #Aerosols/analysis; Bronchial Provocation Tests/instrumentation; Bronchial Provocation Tests/methods; Cyanates/analysis; Humans; Nebulizers and Vaporizers
Tipo

info:eu-repo/semantics/article

article