Eliminating heterophilic antibody interference for ferritin detection using Olympus F(ab')2 based reagent.


Autoria(s): Grouzmann E.; Dayer N.; Bain C.; Neville C.M.; McCusker M.D.; Wang R.Y.
Data(s)

2008

Resumo

Interferences with the Olympus immunoturbidimetric assay for ferritin have been reported because the antibodies used in the immunoassay are derived from rabbits. Rabbits are familiar pets known to be a risk factor for developing heterophilic (or interfering) antibodies. This report shows how the current Olympus Ferritin assay has been improved to eliminate the interference from heterophilic antibodies.

Identificador

http://serval.unil.ch/?id=serval:BIB_6482CB5D4F40

isbn:1433-6510

pmid:19097493

isiid:000261502800005

Idioma(s)

en

Fonte

Clinical Laboratory, vol. 54, no. 9-10, pp. 355-357

Palavras-Chave #Animals; Antibodies, Heterophile/immunology; Antibody Specificity; False Positive Reactions; Ferritins/blood; Ferritins/immunology; Humans; Immunoassay/instrumentation; Immunoassay/methods; Immunoglobulin Fab Fragments/immunology; Nephelometry and Turbidimetry/instrumentation; Nephelometry and Turbidimetry/methods; Rabbits; Sensitivity and Specificity