Interobserver and intraobserver variability of the apparent diffusion coefficient in treated malignant hepatic lesions on a 3.0T machine: measurements in the whole lesion versus in the area with the most restricted diffusion.


Autoria(s): Lu T.L.; Meuli R.A.; Marques-Vidal P.M.; Bize P.; Denys A.; Schmidt S.
Data(s)

2010

Resumo

PURPOSE: To assess the inter/intraobserver variability of apparent diffusion coefficient (ADC) measurements in treated hepatic lesions and to compare ADC measurements in the whole lesion and in the area with the most restricted diffusion (MRDA). MATERIALS AND METHODS: Twenty-five patients with treated malignant liver lesions were examined on a 3.0T machine. After agreeing on the best ADC image, two readers independently measured the ADC values in the whole lesion and in the MRDA. These measurements were repeated 1 month later. The Bland-Altman method, Spearman correlation coefficients, and the Wilcoxon signed-rank test were used to evaluate the measurements. RESULTS: Interobserver variability for ADC measurements in the whole lesion and in the MRDA was 0.17 x 10(-3) mm(2)/s [-0.17, +0.17] and 0.43 x 10(-3) mm(2)/s [-0.45, +0.41], respectively. Intraobserver limits of agreement could be as low as [-0.10, +0.12] 10(-3) mm(2)/s and [-0.20, +0.33] 10(-3) mm(2)/s for measurements in the whole lesion and in the MRDA, respectively. CONCLUSION: A limited variability in ADC measurements does exist, and it should be considered when interpreting ADC values of hepatic malignancies. This is especially true for the measurements of the minimal ADC.

Identificador

http://serval.unil.ch/?id=serval:BIB_6280EC48CD8D

isbn:1522-2586[electronic], 1053-1807[linking]

pmid:20815063

doi:10.1002/jmri.22301

isiid:000281532700015

Idioma(s)

en

Fonte

Journal of Magnetic Resonance Imaging, vol. 32, no. 3, pp. 647-653

Palavras-Chave #liver neoplasms; diffusion weighted imaging; apparent diffusion coefficient; interobserver variability; intraobserver variability; focal liver-lesions; hepatocellular-carcinoma; adc measurements; chemoembolization; reproducibility; breathhold; sequences; abdomen; tumors; mri
Tipo

info:eu-repo/semantics/article

article