Intraoperative assessment of vascular access.


Autoria(s): Saucy F.; Déglise S.; Holzer T.; Salomon L.; Engelberger S.; Corpataux J.M.
Data(s)

2014

Resumo

PURPOSE: The intraoperative quality assessment of the arteriovenous fistula for hemodialysis is an essential process to limit early failure due to technical problems or inadequate vascular quality. This step is not clearly defined in the literature with no recommendations. METHODS: We selected published articles related to the topic of intraoperative quality control of the vascular access for hemodialysis. RESULTS: The intraoperative blood flow measurement greater than 120 ml/min in autologous fistula and less than 320 ml/min in arteriovenous graft was described as predictive factors for early failure. CONCLUSIONS: The blood flow measurement should be performed after the confection of the anastomosis. When blood flow is limited, fistulography is an essential step to assess patency.

Identificador

https://serval.unil.ch/?id=serval:BIB_579552B874D6

isbn:1724-6032 (Electronic)

pmid:24817447

doi:10.5301/jva.5000253

isiid:000342265900002

Idioma(s)

en

Fonte

Journal of Vascular Access, vol. 15, no. Suppl 7, pp. S6-S9

Tipo

info:eu-repo/semantics/article

article