INTRODUCTION: Eddy currents induced by switching of magnetic field gradients can lead to distortions in short echo-time spectroscopy or diffusion weighted imaging. In small bore magnets, such as human head-only systems, minimization of eddy current effects is more demanding because of the proximity of the gradient coil to conducting structures. METHODS: In the present study, the eddy current behavior achievable on a recently installed 7 tesla-68 cm bore head-only magnet was characterized. RESULTS: Residual effects after compensation were shown to be on the same order of magnitude as those measured on two whole body systems (3 and 4.7 T), while using two to three fold increased gradient slewrates.

Identificador

http://serval.unil.ch/?id=serval:BIB_4F3433056F28

isbn:0968-5243

pmid:20012152

doi:10.1007/s10334-009-0192-0

isiid:000274299500004

Idioma(s)

en

Fonte

Magnetic Resonance Materials in Physics Biology and Medicine, vol. 23, no. 1, pp. 39-43