Free-breathing renal magnetic resonance angiography with steady-state free-precession and slab-selective spin inversion combined with radial k-space sampling and water-selective excitation.


Autoria(s): Katoh M.; Spuentrup E.; Stuber M.; Hoogeveen R.; Günther R.W.; Buecker A.
Data(s)

2005

Resumo

The impact of radial k-space sampling and water-selective excitation on a novel navigator-gated cardiac-triggered slab-selective inversion prepared 3D steady-state free-precession (SSFP) renal MR angiography (MRA) sequence was investigated. Renal MRA was performed on a 1.5-T MR system using three inversion prepared SSFP approaches: Cartesian (TR/TE: 5.7/2.8 ms, FA: 85 degrees), radial (TR/TE: 5.5/2.7 ms, FA: 85 degrees) SSFP, and radial SSFP combined with water-selective excitation (TR/TE: 9.9/4.9 ms, FA: 85 degrees). Radial data acquisition lead to significantly reduced motion artifacts (P < 0.05). SNR and CNR were best using Cartesian SSFP (P < 0.05). Vessel sharpness and vessel length were comparable in all sequences. The addition of a water-selective excitation could not improve image quality. In conclusion, radial k-space sampling reduces motion artifacts significantly in slab-selective inversion prepared renal MRA, while SNR and CNR are decreased. The addition of water-selective excitation could not improve the lower CNR in radial scanning.

Identificador

http://serval.unil.ch/?id=serval:BIB_48F6E81893B8

isbn:0740-3194[print], 0740-3194[linking]

pmid:15844141

doi:10.1002/mrm.20467

isiid:000228796900036

Idioma(s)

en

Fonte

Magnetic Resonance In Medicine, vol. 53, no. 5, pp. 1228-1233

Palavras-Chave #Adult; Artifacts; Chi-Square Distribution; Female; Humans; Hypertension, Renovascular/diagnosis; Imaging, Three-Dimensional; Magnetic Resonance Angiography/methods; Male; Renal Artery; Respiration; Statistics, Nonparametric
Tipo

info:eu-repo/semantics/article

article