Preliminary beta spectrum measurements using a magnetic spectrometer.
| Data(s) |
2014
|
|---|---|
| Resumo |
We report the performances of a double focusing magnetic beta spectrometer. The energy resolution was measured using conversion peaks of Cs-137 and Ba-133 at 0.73% for 624 keV, and 1.33% for 124 keV. The counting efficiency as a function of the energy was estimated using a P-32 source and was used to correct the measured spectra of Cs-137. The result was compared with the theoretical spectrum and we found a good agreement. |
| Identificador |
http://serval.unil.ch/?id=serval:BIB_46BC50BE8696 isbn:1872-9800 (Electronic) pmid:24286810 doi:10.1016/j.apradiso.2013.11.022 isiid:000335274000067 |
| Idioma(s) |
en |
| Fonte |
Applied Radiation and Isotopes, vol. 87, pp. 310-314 |
| Tipo |
info:eu-repo/semantics/article article |