Rivaroxaban: Quantification by anti-FXa assay and influence on coagulation tests: a study in 9 Swiss laboratories.


Autoria(s): Asmis L.M.; Alberio L.; Angelillo-Scherrer A.; Korte W.; Mendez A.; Reber G.; Seifert B.; Stricker H.; Tsakiris D.A.; Wuillemin W.A.
Data(s)

2012

Resumo

INTRODUCTION: Rivaroxaban (RXA) is licensed for prophylaxis of venous thromboembolism after major orthopaedic surgery of the lower limbs. Currently, no test to quantify RXA in plasma has been validated in an inter-laboratory setting. Our study had three aims: to assess i) the feasibility of RXA quantification with a commercial anti-FXa assay, ii) its accuracy and precision in an inter-laboratory setting, and iii) the influence of 10mg of RXA on routine coagulation tests. METHODS: The same chromogenic anti-FXa assay (Hyphen BioMed) was used in all participating laboratories. RXA calibrators and sets of blinded probes (aim ii.) were prepared in vitro by spiking normal plasma. The precise RXA content was assessed by high-pressure liquid chromatography-tandem mass spectrometry. For ex-vivo studies (aim iii), plasma samples from 20 healthy volunteers taken before and 2 - 3hours after ingestion of 10mg of RXA were analyzed by participating laboratories. RESULTS: RXA can be assayed chromogenically. Among the participating laboratories, the mean accuracy and the mean coefficient of variation for precision of RXA quantification were 7.0% and 8.8%, respectively. Mean RXA concentration was 114±43μg/L .RXA significantly altered prothrombin time, activated partial thromboplastin time, factor analysis for intrinsic and extrinsic factors. Determinations of thrombin time, fibrinogen, FXIII and D-Dimer levels were not affected. CONCLUSIONS: RXA plasma levels can be quantified accurately and precisely by a chromogenic anti-FXa assay on different coagulometers in different laboratories. Ingestion of 10mg RXA results in significant alterations of both PT- and aPTT-based coagulation assays.

Identificador

http://serval.unil.ch/?id=serval:BIB_45EF032BD7EE

isbn:1879-2472 (Electronic)

pmid:21840043

doi:10.1016/j.thromres.2011.06.031

isiid:000301587400045

Idioma(s)

en

Fonte

Thrombosis Research, vol. 129, no. 4, pp. 492-498

Palavras-Chave #Adult; Anticoagulants/blood; Artifacts; Blood Coagulation Tests/methods; Drug Monitoring/methods; Factor Xa/antagonists & inhibitors; Humans; Male; Morpholines/blood; Reproducibility of Results; Sensitivity and Specificity; Switzerland; Thiophenes/blood
Tipo

info:eu-repo/semantics/article

article