Using a NPWE model observer to assess suitable image quality for a digital mammography quality assurance programme.


Autoria(s): Monnin P.; Bochud F.O.; Verdun F.R.
Data(s)

2010

Resumo

A method of objectively determining imaging performance for a mammography quality assurance programme for digital systems was developed. The method is based on the assessment of the visibility of a spherical microcalcification of 0.2 mm using a quasi-ideal observer model. It requires the assessment of the spatial resolution (modulation transfer function) and the noise power spectra of the systems. The contrast is measured using a 0.2-mm thick Al sheet and Polymethylmethacrylate (PMMA) blocks. The minimal image quality was defined as that giving a target contrast-to-noise ratio (CNR) of 5.4. Several evaluations of this objective method for evaluating image quality in mammography quality assurance programmes have been considered on computed radiography (CR) and digital radiography (DR) mammography systems. The measurement gives a threshold CNR necessary to reach the minimum standard image quality required with regards to the visibility of a 0.2-mm microcalcification. This method may replace the CDMAM image evaluation and simplify the threshold contrast visibility test used in mammography quality.

Identificador

http://serval.unil.ch/?id=serval:BIB_1789ACAE83EE

isbn:1742-3406[electronic], 0144-8420[linking]

pmid:20395413

doi:10.1093/rpd/ncq010

isiid:000277738200087

Idioma(s)

en

Fonte

Radiation Protection Dosimetry, vol. 139, no. 1-3, pp. 459-462