Assessment of the DIXTAL DX-2710 Automated Oscillometric Device for Blood Pressure Measurement with the Validation Protocols of the British Hypertension Society (BHS) and the Association for the Advancement of Medical Instrumentation (AAMI)
Data(s) |
01/12/2002
|
---|---|
Resumo |
OBJECTIVE: To assess the Dixtal DX2710 automated oscillometric device used for blood pressure measurement according to the protocols of the BHS and the AAMI. METHODS: Three blood pressure measurements were taken in 94 patients (53 females 15 to 80 years). The measurements were taken randomly by 2 observers trained to measure blood pressure with a mercury column device connected with an automated device. The device was classified according to the protocols of the BHS and AAMI. RESULT: The mean of blood pressure levels obtained by the observers was 148±38/93±25 mmHg and that obtained with the device was 148±37/89±26 mmHg. Considering the differences between the measurements obtained by the observer and those obtained with the automated device according to the criteria of the BHS, the following classification was adopted: "A" for systolic pressure (69% of the differences < 5; 90% < 10; and 97% < 15 mmHg); and "B" for diastolic pressure (63% of the differences < 5; 83% < 10; and 93% < 15 mmHg). The mean and standard deviation of the differences were 0±6.27 mmHg for systolic pressure and 3.82±6.21 mmHg for diastolic pressure. CONCLUSION: The Dixtal DX2710 device was approved according to the international recommendations. |
Formato |
text/html |
Identificador |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0066-782X2002001500006 |
Idioma(s) |
en |
Publicador |
Sociedade Brasileira de Cardiologia - SBC |
Fonte |
Arquivos Brasileiros de Cardiologia v.79 n.6 2002 |
Palavras-Chave | #validation #automated #oscillometric blood pressure measuring device |
Tipo |
journal article |