Chitosan/AuNPs Modified Graphene Electrochemical Sensor for Label-Free Human Chorionic Gonadotropin Detection


Autoria(s): Teixeira, Sofia; Ferreira, Nádia S.; Conlan, Robert S.; Guy, O. J.; Sales, M. Goreti F.
Data(s)

15/10/2015

15/10/2015

2014

Resumo

A new immunosensor is presented for human chorionic gonadotropin (hCG), made by electrodepositing chitosan/gold-nanoparticles over graphene screen-printed electrode (SPE). The antibody was covalently bound to CS via its Fc-terminal. The assembly was controlled by electrochemical Impedance Spectroscopy (EIS) and followed by Fourier Transformed Infrared (FTIR). The hCG-immunosensor displayed linear response against the logarithm-hCG concentration for 0.1–25 ng/mL with limit of detection of 0.016 ng/mL. High selectivity was observed in blank urine and successful detection of hCG was also achieved in spiked samples of real urine from pregnant woman. The immunosensor showed good detection capability, simplicity of fabrication, low-cost, high sensitivity and selectivity.

Identificador

http://hdl.handle.net/10400.22/6701

10.1002/elan.201400311

Idioma(s)

eng

Publicador

Wiley

Relação

http://onlinelibrary.wiley.com/doi/10.1002/elan.201400311/full

Direitos

openAccess

Palavras-Chave #Immunosensor #Screen-printed electrode #Graphene #Antibody #Chitosan #Human chorionic gonadotropin
Tipo