10 kVp rule - An anthropomorphic pelvis phantom imaging study using a CR system: impact on image quality and effective dose using AEC and manual mode
Data(s) |
13/01/2016
13/01/2016
01/03/2014
|
---|---|
Resumo |
Purpose: This study aims to investigate the influence of tube potential (kVp) variation in relation to perceptual image quality and effective dose for pelvis using automatic exposure control (AEC) and non-AEC in a computed radiography (CR) system. Methods and Materials: To determine the effects of using AEC and non-AEC by applying the 10 kVp rule in two experiments using an anthropomorphic pelvis phantom. Images were acquired using 10 kVp increments (60-120 kVp) for both experiments. The first experiment, based on seven AEC combinations, produced 49 images. The mean mAs from each kVp increment were used as a baseline for the second experiment producing 35 images. A total of 84 images were produced and a panel of 5 experienced observers participated for the image scoring using the 2 AFC visual grading software. PCXMC software was used to estimate the effective dose. Results: A decrease in perceptual image quality as the kVp increases was observed both in non-AEC and AEC experiments, however no significant statistical differences (p> 0.05) were found. Image quality scores from all observers at 10 kVp increments for all mAs values using non-AEC mode demonstrates a better score up to 90 kVp. Effective dose results show a statistical significant decrease (p=0.000) on the 75th quartile from 0.3 mSv at 60 kVp to 0.1 mSv at 120 kVp when applying the 10 kVp rule in non-AEC mode. Conclusion: No significant reduction in perceptual image quality is observed when increasing kVp whilst a marked and significant effective dose reduction is observed. |
Identificador |
Lança L, Franco L, Ahmed A, Harderwijk M, Marti C, Nasir S, et al. 10 kVp rule – An anthropomorphic pelvis phantom imaging study using a CR system: impact on image quality and effective dose using AEC and manual mode. In European Congress of Radiology, Vienna (Austria), 6th to 10th March 2014. Insights Imaging. 2014;5(1 Suppl):S165. http://hdl.handle.net/10400.21/5539 10.1007/s13244-014-0317-5 |
Idioma(s) |
eng |
Publicador |
Springer |
Relação |
http://link.springer.com/article/10.1007/s13244-014-0317-5 |
Direitos |
openAccess http://creativecommons.org/licenses/by-nc/4.0/ |
Palavras-Chave | #Radiology #Pelvis phantom #Effective dose #AEC |
Tipo |
conferenceObject |