Multi-bit sigma-delta modulators with enhanced dynamic-range using non-linear DAC for hearing aids
Data(s) |
12/08/2010
12/08/2010
01/08/2008
|
---|---|
Resumo |
15th IEEE International Conference on Electronics, Circuits and Systems, Malta This paper presents the possibility of employing nonlinear low-resolution DACs in the feedback paths of multi-bit second-order Sigma-Delta modulators. The proposed technique is particularly attractive in applications such as hearing aids, requiring a very large dynamic range and medium signal-tonoise-plus-distortion-ratio. As demonstrated through simulated results in which noise and mismatch effects are included, for the same over-sampling ratio, improvements in the order of 6-to-9 dB in the dynamic range can be achieved when comparing with the same topology employing linear-DACs. |
Identificador | |
Idioma(s) |
eng |
Publicador |
IEEE |
Direitos |
openAccess |
Tipo |
conferenceObject |