Square-wave adsorptive-stripping voltammetric detection in the quality control of fluoxetine
| Data(s) |
26/11/2013
26/11/2013
2007
|
|---|---|
| Resumo |
Electroanalytical methods based on square-wave adsorptive-stripping voltammetry (SWAdSV) and flow-injection analysis with square-wave adsorptive-stripping voltammetric detection (FIA-SWAdSV) were developed for the determination of fluoxetine (FXT). The methods were based on the reduction of FXT at a mercury drop electrode at -1.2 V versus Ag/AgCl, in a phosphate buffer of pH 12.0, and on the possibility of accumulating the compound at the electrode surface. The SWAdSV method was successfully applied in the quantification of FXT in pharmaceutical products, human serum samples, and in drug dissolution studies. Because the presence of dissolved oxygen did not interfere significantly with the analysis, it was possible to quantify FXT in several pharmaceutical products using FIA-SWAdSV. This method enables analysis of up to 120 samples per hour at reduced costs. |
| Identificador |
DOI: 10.1080/00032710701297059 0003-2719 1532-236X |
| Idioma(s) |
eng |
| Publicador |
Taylor & Francis |
| Relação |
Analytical Letters; Vol. 40, Issue 6 http://www.tandfonline.com/doi/abs/10.1080/00032710701297059#.UpTIp0r6Q25 |
| Direitos |
openAccess |
| Palavras-Chave | #Fluoxetine #Square-wave adsorptive-stripping voltammetry #Flow-injection analysis #Electrochemical detection #Drug dissolution |
| Tipo |
article |