Assesing multileaf collimator effect on the build-up region using Monte Carlo method


Autoria(s): Zarza-Moreno, M.; Teixeira, Nuno; Jesus, A. P.; Mora, G.
Data(s)

10/12/2013

10/12/2013

2008

Resumo

Previous Monte Carlo studies have investigated the multileaf collimator (MLC) contribution to the build-up region for fields in which the MLC leaves were fully blocking the openings defined by the collimation jaws. In the present work, we investigate the same effect but for symmetric and asymmetric MLC defined field sizes (2×2, 4×4, 10×10 and 3×7 cm2). A Varian 2100C/D accelerator with 120-leaf MLC is accurately modeled fora6MVphoton beam using the BEAMnrc/EGSnrc code. Our results indicate that particles scattered from accelerator head and MLC are responsible for the increase of about 7% on the surface dose when comparing 2×2 and 10×10 cm2 fields. We found that the MLC contribution to the total build-up dose is about 2% for the 2×2 cm2 field and less than 1% for the largest fields.

Identificador

Zarza-Moreno M, Teixeira N, Jesus AP, Mora G. Assesing multileaf collimator effect on the build-up region using Monte Carlo method. Pol J Med Phys Eng. 2008;14(3):163-82.

1425-4689

http://hdl.handle.net/10400.21/2987

Idioma(s)

eng

Relação

http://www.degruyter.com/view/j/pjmpe.2008.14.issue-3/v10013-008-0014-0/v10013-008-0014-0.xml?format=INT

Direitos

openAccess

Palavras-Chave #Physics #Surface dose #Build-up dose #MLC #IMRT #Monte Carlo
Tipo