On a possible mechanism for peripheral nerve stimulation during magnetic resonance imaging scans


Autoria(s): Forbes, Lawrence K.; Crozier, Stuart
Data(s)

01/02/2001

Resumo

When patients undergo a magnetic resonance imaging scan, they are subject to both strong static and temporal magnetic fields. The temporal fields are designed to vary at each point in the region being imaged. This is achieved by the use of gradient coils. However, when the gradient coils are switched very rapidly, the strongly time-varying magnetic fields produced can be responsible for stimulating nerves in the peripheral regions of the body. This paper gives a somewhat novel explanation for this phenomenon. The physical mechanism suggested is supported by an illustrative theoretical calculation.

Identificador

http://espace.library.uq.edu.au/view/UQ:60978

Idioma(s)

eng

Publicador

IOP Publishing Ltd

Palavras-Chave #Engineering, Biomedical #Radiology, Nuclear Medicine & Medical Imaging #Field Stimulation #Gradient #Neurons #Coils #C1 #090399 Biomedical Engineering not elsewhere classified