Validation of a manual headspace gas chromatography method for determining volatile compounds in biological fluids
Contribuinte(s) |
UNIVERSIDADE DE SÃO PAULO |
---|---|
Data(s) |
19/10/2012
19/10/2012
2008
|
Resumo |
Background: We report the validation of a method for the determination of acetaldehyde, acetone, methanol, and ethanol in biological fluids using manual headspace sample introduction and an acetonitrile internal standard. Method: This method uses a capillary column (I = 30 m, I.D. = 0.25 mm, dF = 0.25 mu m) installed in a gas chromatography-flame ionization detector (GC-FID) apparatus with a run time of 7.5 minutes. Results: Analysis of the retention times and the resolution of the analyte peaks demonstrated excellent separation without widening of the peaks. Precision and accuracy were good (interassay precision < 15% and recovery between 85% and 115%) in both blood and urine. Conclusion: The method was linear (r > 0.09) over the analytical measurement range (AMR) of each analyte. |
Identificador |
LABMEDICINE, v.39, n.1, p.42-45, 2008 0007-5027 http://producao.usp.br/handle/BDPI/24077 10.1309/EA876C1PVC0UXPH0 |
Idioma(s) |
eng |
Publicador |
AMER SOC CLINICAL PATHOLOGY |
Relação |
Labmedicine |
Direitos |
restrictedAccess Copyright AMER SOC CLINICAL PATHOLOGY |
Palavras-Chave | #SOLID-PHASE MICROEXTRACTION #ALCOHOL DETERMINATION #MASS-SPECTROMETRY #ETHANOL #BLOOD #Medical Laboratory Technology |
Tipo |