Safety Profile of Trocar and Insufflation Needle Access Systems in Laparoscopic Surgery


Autoria(s): PASSEROTTI, Carlo C.; BEGG, Nikolai; PENNA, Frank J.; PASSEROTTI, Ana Maria A. M. S.; LEITE, Katia R. M.; ANTUNES, Alberto Azoubel; Srougi, Miguel; RETIK, Alan B.; NGUYEN, Hiep T.
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

19/10/2012

19/10/2012

2009

Resumo

BACKGROUND: The most common laparoscopic complications are associated with trocar insertion. The purpose of this study was to develop an objective method of evaluating the safety profile of various access devices used in laparoscopic surgery. STUDY DESIGN: In 20 swine, 6 bladed and 2 needle access devices were evaluated. A force profile was determined by measuring the force required to drive the trocar or needle through the fascia and into the peritoneum, at 0 and 10 mmHg. The amount Of tissue deformation, the length of blade exposed, and the duration of exposure were measured using a high-speed digital imaging system. RESULTS: The needle system without the sheath required the least driving force and had the most favorable force profile. In contrast, the bladed, nonretractable trocar system required a higher driving force and a rapid loss of resistance. Insertion under a pneumoperitoneum did not significantly alter the force profile of the various access devices except for the amount of tissue deformation. With the bladed system, the blade itself was exposed for an average of 0.5 to 1.0 seconds for a distance of 4.5 to 5.0 cm. In comparison, the needle system was exposed for 0.2 seconds for a distance of 1.8 cm. CONCLUSIONS: We developed a reproducible method of measuring the forces required to place the access systems, their pattern of resistance loss, and the characteristics of the blade exposure. These parameters may provide an adjunctive and objective measurement of safety, allowing for more direct comparison between various trocar designs. (J Am Coll Surg 2009;209:222-232. (C) 2009 by the American College of Surgeons)

CAPES

FAPESP

Brazilian Funding Agencies (FINEP)

Identificador

JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, v.209, n.2, p.222-232, 2009

1072-7515

http://producao.usp.br/handle/BDPI/23257

10.1016/j.jamcollsurg.2009.03.022

http://dx.doi.org/10.1016/j.jamcollsurg.2009.03.022

Idioma(s)

eng

Publicador

ELSEVIER SCIENCE INC

Relação

Journal of the American College of Surgeons

Direitos

restrictedAccess

Copyright ELSEVIER SCIENCE INC

Palavras-Chave #VERESS NEEDLE #CHOLECYSTECTOMY #COMPLICATIONS #Surgery
Tipo

article

original article

publishedVersion