Direct visualisation of B-1 inhomogeneity by flip angle dependency


Autoria(s): Zelaya, FO; Roffmann, WU; Crozier, S; Teed, S; Gross, D; Doddrell, DM
Data(s)

01/01/1997

Resumo

Inhomogeneities in the spatial distribution of the excitatory Radio Frequency (RF) field, are still a dominant source of artifacts and loss of signal to noise ratio in MR imaging experiments, A number of strategies have been proposed to quantify this distribution, However, in this technical note we present a relatively simple MR imaging procedure which can be used to visualise RF inhomogeneities directly either by means of the magnitude or the phase of an image. To visualise the RF field distribution in both the inner and outer volumes of the coil, we have performed experiments in which the entire coil is submerged in a non-conducting fluid, To the best of our knowledge this strategy has not been used previously in order to evaluate coil performance, Finally, we demonstrate that the method is sensitive enough to reveal the effects of the sample properties on the effective RF wavelength of the transmitted field. (C) 1997 Elsevier Science Inc.

Identificador

http://espace.library.uq.edu.au/view/UQ:57770

Idioma(s)

eng

Palavras-Chave #Radiology, Nuclear Medicine & Medical Imaging #Rf Inhomogeneity #Flip Angle Error #Birdcage Resonator #Nmr #Coil
Tipo

Journal Article