Ventilation Patterns Influence Airway Secretion Movement


Autoria(s): VOLPE, Marcia S.; ADAMS, Alexander B.; AMATO, Marcelo B. P.; MARINI, John J.
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

19/10/2012

19/10/2012

2008

Resumo

BACKGROUND: Retention of airway secretions is a common and serious problem in ventilated patients. Treating or avoiding secretion retention with mucus thinning, patient-positioning, airway suctioning, or chest or airway vibration or percussion may provide short-term benefit. METHODS: In a series of laboratory experiments with a test-lung system we examined the role of ventilator settings and lung-impedance on secretion retention and expulsion. Known quantities of a synthetic dye-stained mucus simulant with clinically relevant properties were injected into a transparent tube the diameter of an adult trachea and exposed to various mechanical-ventilation conditions. Mucus-simulant movement was measured with a photodensitometric technique and examined with image-analysis software. We tested 2 mucus-simulant viscosities and various peak flows, inspiratory/ expiratory flow ratios, intrinsic positive end-expiratory pressures, ventilation waveforms, and impedance values. RESULTS: Ventilator settings that produced flow bias had a major effect on mucus movement. Expiratory How bias associated with intrinsic positive end-expiratory pressure generated by elevated minute ventilation moved mucus toward the airway opening, whereas intrinsic positive end-expiratory pressure generated by increased airway resistance moved the mucus toward the lungs. Inter-lung transfer of mucus simulant occurred rapidly across the ""carinal divider"" between interconnected test lungs set to radically different compliances; the mucus moved out of the low-compliance lung and into the high-compliance lung. CONCLUSIONS: The movement of mucus simulant was influenced by the ventilation pattern and lung impedance. Flow bias obtained with ventilator settings may clear or embed mucus during mechanical ventilation.

Identificador

RESPIRATORY CARE, v.53, n.10, p.1287-1294, 2008

0020-1324

http://producao.usp.br/handle/BDPI/21314

http://apps.isiknowledge.com/InboundService.do?Func=Frame&product=WOS&action=retrieve&SrcApp=EndNote&UT=000259988900006&Init=Yes&SrcAuth=ResearchSoft&mode=FullRecord

Idioma(s)

eng

Publicador

DAEDALUS ENTERPRISES INC

Relação

Respiratory Care

Direitos

restrictedAccess

Copyright DAEDALUS ENTERPRISES INC

Palavras-Chave #airway clearance #mechanical ventilation #mucus #secretions #MANUAL HYPERINFLATION #BRONCHIAL-SECRETIONS #GOLD STAGES #CLEARANCE #MUCUS #PULMONARY #INTUBATION #TRANSPORT #SYMPTOMS #REMOVAL #Critical Care Medicine #Respiratory System
Tipo

article

original article

publishedVersion