Quantification of carbamazepine, carbamazepine-10,11-epoxide, phenytoin and phenobarbital in plasma samples by stir bar-sorptive extraction and liquid chromatography


Autoria(s): QUEIROZ, Regina Helena Costa; BERTUCCI, Carlo; MALFARA, Wilson Roberto; DREOSSI, Sonia Aparecida Carvalho; CHAVES, Andrea Rodrigues; VALERIO, Daniel Augusto Rodrigues; QUEIROZ, Maria Eugenia Costa
Contribuinte(s)

UNIVERSIDADE DE SÃO PAULO

Data(s)

19/10/2012

19/10/2012

2008

Resumo

A sensitive and reproducible stir bar-sorptive extraction and high-performance liquid chromatography-UV detection (SBSE/HPLC-UV) method for therapeutic drug monitoring of carbamazepine, carbamazepine-10,11-epoxide, phenytoin and phenobarbital in plasma samples is described and compared with a liquid:liquid extraction (LLE/HPLC-UV) method. Important factors in the optimization of SBSE efficiency such as pH, extraction time and desorption conditions (solvents, mode magnetic stir, mode ultrasonic stir, time and number of steps) assured recoveries ranging from 72 to 86%, except for phenytoin (62%). Separation was obtained using a reverse phase C-18 column with UV detection (210 nm). The mobile phase consisted of water: acetonitrile (78:22, v/v). The SBSE/HPLC-UV method was linear over a working range of 0.08-40.0 mu g mL(-1) for carbamazepine, carbamazepine-10,11-epoxide and phenobarbital and 0.125-40.0 mu g mL(-1) for phenytoin, The intra-assay and inter-assay precision and accuracy were studied at three concentrations (1.0, 4.0 and 20.0 mu g mL(-1)). The intra-assay coefficients of variation (CVs) for all compounds were less than 8.8% and all inter-CVs were less than 10%. Limits of quantification were 0.08 mu g mL(-1) for carbamazepine, carbamazepine-10,11-epoxide and phenobarbital and 0.125 mu g mL(-1) for phenytoin. No interference of the drugs normally associated with antiepileptic drugs was observed. Based on figures of merit results, the SBSE/HPLC-UV proved adequate for antiepileptic drugs analyses from therapeutic levels. This method was successfully applied to the analysis of real samples and was as effective as the LLE/HPLC-UV method. (c) 2008 Elsevier B.V. All rights reserved.

Identificador

JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, v.48, n.2, Special Issue, p.428-434, 2008

0731-7085

http://producao.usp.br/handle/BDPI/20735

10.1016/j.jpba.2008.03.020

http://dx.doi.org/10.1016/j.jpba.2008.03.020

Idioma(s)

eng

Publicador

PERGAMON-ELSEVIER SCIENCE LTD

Relação

Journal of Pharmaceutical and Biomedical Analysis

Direitos

restrictedAccess

Copyright PERGAMON-ELSEVIER SCIENCE LTD

Palavras-Chave #stir bar-sorptive extraction #high-performance liquid chromatography #carbamazepine #carbamazepine-10,11-epoxide #phenytoin #phenobarbital #therapeutic drug monitoring #SOLID-PHASE MICROEXTRACTION #MASS-SPECTROMETRY #THERMAL-DESORPTION #ANTIEPILEPTIC DRUGS #DRINKING-WATER #HUMAN SERUM #METABOLITE #4-NONYLPHENOL #URINE #4-TERT-OCTYLPHENOL #Chemistry, Analytical #Pharmacology & Pharmacy
Tipo

article

proceedings paper

publishedVersion